Skip to content

Processing Parameters

All analysis settings in CaloCem live in one object, ProcessingParameters. You create it once, change the values you need and pass it to the methods of Measurement:

from calocem import Measurement, ProcessingParameters

processparams = ProcessingParameters()
processparams.cutoff.cutoff_min = 30
processparams.spline_interpolation.apply = True

tam = Measurement(folder="calo_data", processparams=processparams)
peaks = tam.get_peaks(processparams, show_plot=False)

If a method is called without processparams, it uses the parameters the Measurement was created with (or the defaults).

The parameters are grouped by topic. Each group is a small dataclass, so you can also create a group directly, e.g. ProcessingParameters(cutoff=CutOffParameters(cutoff_min=30)).

cutoff

The beginning of a measurement is usually disturbed by mixing and thermal equilibration. cutoff_min removes it from the analysis.

Field Default Description
cutoff_min 60 Time in minutes. Data with time_s < cutoff_min * 60 is ignored. 0 disables the cutoff.

cutoff_min is honoured by peak detection, dormant period, maximum and mean slope, first ascending slope, deconvolution and the related plots.

Exceptions

  • get_cumulative_heat_at_hours does not discard data; instead, the heat released up to cutoff_min is subtracted from the heat at target_h.
  • get_peak_onsets ignores cutoff_min and uses its own time_discarded_s argument.
  • The half-maximum search of get_astm_c1679_characteristics uses the unfiltered data; only the peak detection it relies on honours the cutoff.

time_constants

Time constants of the calorimeter for the Tian correction, in seconds. Used by apply_tian_correction only.

Field Default Description
tau1 300 First time constant.
tau2 100 Second time constant. Set to None to correct with tau1 only.

peakdetection

Passed to scipy.signal.find_peaks when searching peaks in the heat flow.

Field Default Description
prominence 1e-5 Minimum peak prominence, in the unit of the analysed column (W/g). Decrease it to find smaller peaks.
distance 100 Minimum distance between peaks as a number of data points, not a time.

Peak detection is the first step of get_peaks, get_dormant_period_heatflow, get_astm_c1679_characteristics and get_mainpeak_params.

Smoothing

Derivatives of measured heat flow are noisy. Three groups control smoothing. They act at different points, so it matters which analysis you run:

Group Fields (default) What is smoothed
rolling_mean apply (False), window (11 points) Heat flow, before the gradient is taken. Used by the maximum slope analysis and plot_heatflow_with_gradient.
median_filter apply (False), size (7 points) The first and second derivative of the heat flow (not the heat flow itself).
spline_interpolation apply (False), smoothing_1st_deriv (1e-9), smoothing_2nd_deriv (1e-9) A cubic smoothing spline is fitted to the first and second derivative. The values are passed as s to UnivariateSpline; larger values mean smoother curves.

The smoothed derivatives are used by the maximum slope analysis (get_maximum_slope, get_mainpeak_params, get_peak_onset_via_max_slope), plot_heatflow_with_gradient and apply_tian_correction. The mean slope, first ascending slope, peak detection and deconvolution work on the unsmoothed data.

downsample

Large files can be thinned out when the Measurement is created. Points are kept preferentially where the heat flow curve bends, so peaks stay sharp while flat regions are thinned out. Values are picked from the measured data, not interpolated.

Field Default Description
apply False Enable downsampling.
num_points 1000 Target number of points per sample (an upper bound).
smoothing_factor 1e-6 Smoothing of the spline used to estimate the curvature.
baseline_weight 0.1 Share of points spread evenly, so flat regions are not emptied.
section_split False Downsample the data before and after section_split_time_s separately.
section_split_time_s 1000 Split time in seconds.

Downsampling only takes effect when passed to the constructor: Measurement(folder, processparams=processparams). Save the result with tam.save(...) to avoid repeating it, see Loading data.

slope_analysis

Controls the analyses of the main hydration peak, see Main Peak Parameters.

Field Default Description
flank_fraction_start 0.35 Start of the ascending flank, as fraction of the height between the minimum before the peak and the peak.
flank_fraction_end 0.55 End of the ascending flank, same scale.
window_size 0.1 Width of the averaging window for the mean slope, as fraction of the time span of the flank.
first_ascending_fraction_of_max 0.2 method="ascending": heat flow threshold as fraction of the maximum (after cutoff). Must be in (0, 1].
first_ascending_range_method "fraction" method="ascending": "fraction" uses the flank fractions above, "delta" uses a fixed heat-flow rise.
first_ascending_delta_y_w_g 0.001 "delta" mode: rise of the heat flow in W/g that defines the segment.
flexible 0.0 "delta" mode: widens the delta for high peaks; 0 disables it.

deconvolution

Used by get_deconvolution and get_left_peak_inflection_tangent_intersection, which fit a sum of peaks plus a baseline to the heat flow.

Field Default Description
n_peaks 2 Number of peaks to fit.
baseline_mode "constant" "constant", "linear", "chebyshev" or "none".
chebyshev_degree 2 Polynomial degree for baseline_mode="chebyshev".
max_nfev 5000 Maximum number of optimizer iterations.
max_fit_points 2500 The data is subsampled to at most this many points for the fit.
min_points 30 Samples with fewer points (after cutoff) are skipped.
min_peak_distance_fraction 0.05 Minimum distance of the initial peak guesses, as fraction of the number of points.
min_sigma_fraction, max_sigma_fraction 0.005, 0.3 Width bounds of Gaussian peaks, as fraction of the time range.
min_peak_time_separation_fraction 1e-4 Minimum separation of the fitted peak times, as fraction of the time range.
relative_intensity_upper_bounds None Optional list (one value per peak, each in (0, 1]) capping each peak's share of the total amplitude.
peak_width_upper_bounds None Optional list (one value per peak) of maximum peak widths.

The peak shape is selected with the peak_shape argument of get_deconvolution ("lognormal" or "gaussian"); the peak_shape field is currently ignored.

plotting

Used by the analysis plots of get_mainpeak_params (and the methods built on it) and, for figsize, by plot_heatflow_with_gradient. Measurement.plot does not use these settings.

Field Default Description
figsize (10, 6) Figure size in inches, used when no ax is passed.
time_unit "seconds" "seconds", "minutes" or "hours".
heat_unit "W" "W" or "mW".
show_plot_title True Show a title above each plot.
plot_title "filename" With metadata loaded: a list of metadata column names whose values form the title, e.g. ["cement_name", "w_c"].
legend_pos "best" "best" or "outside" (right of the plot).
xlims, ylims None Axis limits as (min, max) in plot units. ylims is only applied together with xlims.

Sample-specific parameters

Different samples sometimes need different settings. Rules map a regular expression on the sample name to parameter overrides:

processparams.add_sample_param_rule(
    r"^c3a",  # samples whose name starts with "c3a"
    {"slope_analysis": {"first_ascending_range_method": "delta"}},
)

Rules with a higher priority are applied later and therefore win.

Warning

Sample rules are currently only honoured by get_mainpeak_params(method="ascending"). All other analyses use the base parameters for every sample.

Reserved parameters

The groups gradient_peakdetection, nonlin_savgol and preprocess and the field cutoff.cutoff_max exist but are not used by any analysis yet. Changing them has no effect.